-
On the gear side of the cartridge, remove the four
screws on the end cap. See Figure 3
-
Carefully remove the end cap. Note the location of
the gears. See Figures 4 & 5
-
From the fill plug side, remove the
three screws. See Figure 6
FIGURE 3 |
FIGURE 4 |
|
|
FIGURE 5 |
FIGURE 6 |
-
Carefully remove the supply chamber. Be
careful not to damage the drum! See Figure 7
-
On the fill plug side end cap there is a
second smaller end cap, or contact plate. This plate is held in with four very
small screws. Use a jewelers Phillips head screwdriver to remove them. See
Figure 8
FIGURE 7 |
FIGURE 8 |
-
Remove the contact plate. A gear may
come loose. Be sure not to lose it. (It may stick to the inside of the plate).
See Figure 9
-
Remove the large end cap, and drum
cover. See Figure 10
FIGURE 9 |
FIGURE 10 |
-
The PCR will be left sitting next to the
drum. Remove the drum and PCR, place aside. Do not touch the PCR with your
hands, as the oil normally present in your skin will contaminate the roller. See
Figure 11
The PCR in these cartridges is a felt type
that should only be cleaned with clean, dry compressed air. Do not use any
chemicals on it, as there is a conductive coating on the felt that will be
removed. The PCR’s in these cartridges will typically last for up to two cycles.
New replacement PCR’s will probably never be available.
-
Remove the wiper blade and two screws.
Leave the two small metal plates on the wiper blade. Believe it or not they are
the contacts for the PCR! When new wiper blades become available, these contacts
should be swapped over. See Figures 12 & 13
-
Clean out the waste chamber. Note that
there is an auger that pushes the waste toner back into the supply chamber. Make
sure that the auger turns freely, and that the drive gear on the outside of the
hopper does not get lost.
-
Coat the wiper blade with your
preferred lubricant, and install in the cartridge. Install the two screws. Align
the blade with the three tabs, and drop in place. See Figure 14
FIGURE 11 |
FIGURE 12 |
|
|
FIGURE 13 |
FIGURE 14 |
-
Clean the PCR contacts with a cotton
swab and alcohol. See Figure 15
-
Remove the fill plug from the supply
chamber. Dump out any remaining toner, and vacuum clean. See Figure 16
FIGURE 15 |
FIGURE 16 |
Although at this stage it is tempting to
just fill the supply, and close it up, it is very important that this not be
done. The Doctor Blade in these cartridges tends to get a buildup on it that
will interfere with toner distribution on the developer roller, causing
streaking down the page. If not cleaned, this can happen in as soon as 1000
pages. This blade needs to be cleaned every cycle.
-
Remove the two small springs on the
developer roller tensioners with the spring hook. See Figure 17.
-
Turn both of the developer roller clips
up so that they unlock, and remove. Both clips have been colorized for ease of
identification. See Figures 18 & 19
-
Remove the developer roller. Be careful
when handling the roller as it has a sleeve around it that is easily damaged. We
have found it best to handle the roller by the metal shaft only. Do not touch
the sleeve with your skin. See Figure 20.
FIGURE 17 |
FIGURE 18 |
|
|
FIGURE 19 |
FIGURE 20 |
-
Remove the two screws from the doctor
blade. See Figure 21.
-
Remove the doctor blade being very
careful not to damage the seal foam. It may stick to the blade as it is removed.
Separate the foam from the blade, and remove the blade. See Figure 22.
-
Clean out all the remaining toner from
the hopper. Be especially careful to clean any toner residue off the metal feed
roller. It is also important not to damage the clear plastic tensioners on
either side of the hopper. See Figures 23 & 24
FIGURE 21 |
FIGURE 22 |
|
|
FIGURE 23 |
FIGURE 24 |
-
Place the doctor blade foam back in its
proper position. See Figure 25
-
Clean the doctor blade with Acetone
followed by alcohol. Make sure it is dry, and install in the cartridge. Make
sure that the Dr. blade contact is behind the blade, and that the tensioners are
on top. See Figures 26 & 27
These cartridges come new with out seals.
The supply hopper is tight enough not to leak (much). Actually they hold up
pretty good. Just make sure they are packed good if shipping them.
-
Install the developer roller. Make sure
that the spacers on each side of the roller are straight, and in their proper
place. See Figure 28
FIGURE 25 |
FIGURE 26 |
|
|
FIGURE 27 |
FIGURE 28 |
-
Install both developer roller clips,
larger on the gear side, smaller on the fill plug side. Lock both clips in
place. Make sure that the clear tensioner strips are under the clip tabs. See
Figures 29, 30, & 31.
-
Install the two small springs on the tensioners. Make sure you keep the hook side of the spring down! See Figure 32
FIGURE 29 |
FIGURE 30 |
|
|
FIGURE 31 |
FIGURE 32 |
-
Fill with 300g 151 toner. See Figure 33
-
Replace the fill plug, check for leaks.
See Figure 34
-
Check the geared end cap to see if the
grease is contaminated with toner. If it is, clean it all off with alcohol, and
replace it with white lithium grease. Be sure to try and put the same amount
back that was removed. See Figure 35
Install the right side end cap (Has two
gears mounted on it), onto the waste hopper. See Figure 36
FIGURE 33 |
FIGURE 34 |
|
|
FIGURE 35 |
FIGURE 36 |
-
Install two of the screws as indicated.
See Figure 37
-
Install the cleaned (Blown off) PCR
into the end cap. The thick shaft side fits into the end cap. Do not touch the
PCR with your hands, as the oil normally present in your skin will contaminate
the roller. Make sure that the gears are aligned. See Figure 38
FIGURE 37 |
FIGURE 38 |
-
Install the drum large gear side to the
right. See Figure 39
If the OEM drum is damaged, new drums are
available with out gears, the gears must be swapped out.
-
Install the toner hopper into the end
cap. It should just snap in place. Make sure that the cut gear shaft fits into
the hopper properly. You may have to turn the gears slightly to align them.
Install the two remaining screws. See Figures 40 & 41
-
Remove the drum cover from the left end
cap. Turn it up so that it can be gently pried from its holder. See Figure 42
FIGURE 39 |
FIGURE 40 |
|
|
FIGURE 41 |
FIGURE 42 |
-
Place the loose gear from the end cap
by the PCR gear. See Figure 43
-
Install the end cap onto the cartridge,
install the two screws as shown. See Figure 44
-
This is tricky, the shaft from the
contact plate has to go through the end cap and the loose gear. The metal drum
axle must fit into the drum and must also line up with the two hoppers. Don’t
force it; keep playing with it until it fits. See Figure 45
-
Install the four small screws in the
contact plate. See Figure 46
FIGURE 43 |
FIGURE 44 |
|
|
FIGURE 45 |
FIGURE 46 |
-
Install the three remaining large
screws. See Figure 47
-
Install the drum cover back on to the
end cap. Make sure that the spring is set properly, and that the end of the
cover fits into it’s slot on the opposite side. See Figures 48 & 49.
The drum does not drive the toner supply
section of the cartridge, only the waste section. The toner supply is driven
separately by the large gear on the bottom of the toner hopper. See Figure 50.
FIGURE 47 |
FIGURE 48 |
|
|
FIGURE 49 |
FIGURE 50 |
So if you normally spin the drum a few
times to ensure that all the gears are aligned, and that the drum & wiper blade
are lubricated, don’t panic! Only the drum is supposed to turn!
The cartridge is finished!
This last part of re-assembling the cartridge is one of
the worst I have seen. It must be a nightmare for the production line when they
build these! As the commercials say; “From the minds of Minolta”!